Healthport® Spinal

Fully implantable spinal and epidural access system

 

Characteristics of the chamber

  • Small dimensions and low weight with very large septum surface to improve access and duration of use
  • Interior of the chamber entirely of titanium with the highest bio- and chemo compatibility
  • Internal filters: 50 μ
  • Suitable for MRI and CT scans


Characteristics of the catheter

  • Made of polyurethane
  • Length marking in cm in the distal 20 cm
  • Open-ended with rounded tip or closed with three lateral openings
  • Radiopaque

 

LP Titan Spinal

Full metal chamber, length marking in cm and measurements every 5 cm

CatheterInt Ø mm /
Ext Ø mm
GaugeLength in
mm
Volume
ml
TipOrder code
Polyurethane 0.5 / 1.0 19 900 0.18 open 201.15.10.61
Polyurethane 0.5 / 1.0 19 900 0.18 closed 201.15.10.71
Chamber: Full titanium, base diameter: 26 mm, height: 11.5 mm, volume: 0.53 ml, weight: 7.5 g
Septum: silicone, diameter: 11.8 mm, maximum number of punctures: 2000 with 22G /1000 with 19G

 

ETI Titanium spinal

Full metal chamber, length marking in cm and measurements every 5 cm, easy placement with a small incision

CatheterInt Ø mm /
Ext Ø mm
GaugeLength in
mm
Volume
ml
TipOrder code
Polyurethane 0.5 / 1.0 19 900 0.18 open 201.17.10.61
Polyurethane 0.5 / 1.0 19 900 0.18 closed 201.17.10.71
Chamber: Full titanium, base length: 27.5 mm, base width: 19 mm, height: 11.5 mm, volume: 0.53 ml
Septum: silicone, diameter: 12.8 mm, maximum number of punctures: 2000 with 22G /1000 with 19G


Package includes: titanium chamber, catheters, screw connector, 1x Tuohy needle 17G x 90 mm, 1x tunneler steel,
Huber cannula 25G 24.5mm, 1x blunt cannula

KONTAKT PONTIFEX MEDICAL

Fabriksvænget 18 B // DK-4130 Viby Sjælland

Tel. : + 45 61 39 37 60
hd@pontifexmedical.com

+ 45 61 39 37 60

KONTAKT PROMECON GMBH

Admiralitätstraße 60 // 20459 Hamburg

Tel. : +49 40 3690 1690
Fax : +49 40 3690 1699
info@promecon-medical.de