Accufuser
Accufuser is a single-use elastomeric pump that works using the elasticity principle with a balloon reservoir made of medical-grade silicone inside a protective casing and a flow restrictor made of high-quality boron glass. Accufuser makes possible the highest possible mobility for the patients during their infusion and is a secure alternative to electronic pump systems.
Its advantages are:
- Easy handling for patients, doctors, nursing staff, and pharmacists.
- Small and discreet for patients
- No irritating alarms
- Works without batteries or electricity
- Safety due to closed single-use system, kink-resistant tubing, and high-quality flow restrictor
- Large selection of infusion rates and filling volumes for optimal adaptation to therapeutic needs
Article list 100 ml
Accufuser continuous 100 ml nominal filling volume. Recommended maximum filling volume 110 ml, recommended minimum filling volume not below 60 ml.
Code | Flow rate | Infusion time in days | Infusion time in hours | PZN | Aid number |
---|---|---|---|---|---|
C0130M | 13 ml/h | 7.5 h | 5950143 | 03 99 04 0 900 | |
C0050M | 5.0 ml/h | 1 day | 22 h | 5950114 | 03 99 04 0 043 |
C0040M | 4.0 ml/h | 1 day | 25 h | 5950083 | 03 99 04 0 900 |
C0020M | 2.0 ml/h | 2 days | 50 h | 5950077 | 03 99 04 0 042 |
C0010M | 1.0 ml/h | 3-4 days | 100 h | 5950060 | 03 99 04 0 041 |
C0005M | 0.5 ml/h | 5-9 days | 200 h | 5950048 | 03 99 04 0 040 |
The infusion time given a flow rate (ml/h) is calculated using the filling volume. Infusion time (h) is equal to filling volume (ml) divided by flow rate (ml/h). E.g. a pump with 4 ml/h flow rate has to be filled to 96 ml to obtain a flow time of one day. if a reduced flow time is required, a flow rate of 5 ml/h and a pump filled with 110ml is required.
Article list 300 ml
Accufuser continuous 300 ml nominal filling volume. Recommended maximum filling volume 330 ml, recommended minimum filling volume not below 180 ml.
Code | Flow rate | Infusion time in days | Infusion time in hours | PZN | Aid number |
---|---|---|---|---|---|
C0100L | 10 ml/h | 1 day | 30 h | 7797119 | 03 99 04 0 049 |
C0080L | 8 ml/h | 1½ days | 37.5 h | 7770617 | 03 99 04 0 048 |
C0050L | 5.0 ml/h | 2 - 2½ days | 60 h | 7770600 | 03 99 04 0 047 |
C0040L | 4.0 ml/h | 3 days | 75 h | 03 99 04 0 900 | |
C0020L | 2.0 ml/h | 4-6 days | 150 h | 7703934 | 03 99 04 0 046 |
C0015L | 1.5 ml/h | 7-9 days | 200 h | 7704419 | 03 99 04 0 045 |
C0010L | 1.0 ml/h | 10 Days | 300 h | 5950195 | 03 99 04 0 900 |
The infusion time given a flow rate (ml/h) is calculated using the filling volume. Infusion time (h) is equal to filling volume (ml) divided by flow rate (ml/h). E.g. a pump with 2 ml/h flow rate has to be filled to 240 ml to obtain a flow time of five day.
Further models upon request