Seesite Port for high pressure infusion

 

 

Serie

Reference

Pre- connected
references

Catheter

Outer ⌀ (mm)

Inner ⌀ (mm)

Introducer
size

Port dimensions (LxWxH) / and weight

Internal volume (ml)

Maximum flow rate recommended with 25 cm length Catheter

Viskosity (cP) / Maximum concentration of lodine recommended for contrast media

Micro

2005SEE

 

Silicone

1.65 (5F)

0.65

5F

22 x 17 x 9 mm

2.9 g

0.2

1 ml/s

6 cP or 300mgl/ml (e.g.: Xenetix®, 300, pre-heated at 37°C)*

2105SEE

x

Mini

3007SEE

  Silicone

2.16 (6.5F)

1.02

7F

26 x 21

x 10 mm

5.5 g

0.4

3 ml/s

10 cP or 350mgl/ml (e.g.: Xenetix® 350, pre-heated at 37°C)*

3107SEE

x

3017SEE

 

Polyurethane

2.30 (7.0F)

1.45

5 ml/s

3117SEE

x

3008SEE

 

Silicone

2.40 (7.2F)

1.20

 

8F

3108SEE

x

Standard and High flow

4008SEE

 

Silicone

2.40 (7.2F)

1.20

31 x 22

x 12 mm

7.7 g

0.6

4108SEE

x

4018SEE

 

Polyurethane

2.65 (8.0F)

1.70

4118SEE

x

4019SEE

 

Polyurethane

3.00 (9.0F)

1.90

9F

4119SEE

x

40010SEE

 

Silicone

3.18 (9.5F)

1.57

10F

41010SEE

x

*Xenetix® is a registered trade mark of Guerbet Laboratories.

For all references the maximum huber needle diameter recommended for contrast media injection: 19G or 20G exept for 2005SEE and 2105SEE references. For these two references the Maximum Huber needle diameter recommended for contrast media injection is 20G or 22G. (4)

KIT INCLUDES

  • 1 hybrid port (Polyoxymethylene and titanium, septum in silicone )
  • 1 radiopaque silicone or polyurethane catheter of 60cm (centimeter marking starts at 5cm and ends at 40cm)
  • 2 connecting rings, radiopaque (only one for pre-connected references)
  • 1 straight Huber needle (22G or 23G**)
  • 1 vein pick
  • 1 flushing connector
  • 1 echogenic introducer needle (18G or 20G**)
  • 1 peel-away introducer sheet (12 cm** or 17 cm)
  • 1 J guidewire marked every 10 cm (0.018"/40 cm** or 0.035"/60 cm)
  • 1 tunneling device (18 cm** or 23 cm)
  • 1 10ml syringe
  • 1 Raulerson bulb

ALSO INCLUDED

  • 1 safety Huber needle compatible with high pressure injection of contrast media, PPS® CT (20G-20mm ref.802009 or 22G**- 20mm ref.80200)
  • 1 set dedicated to ultrasound guided venipuncture:
  • 1 Sterile dressing
  • Latex free protective sheath for ultrasound probe
  • 2 Sterile elastics
  • Sterile gel

(1) Ring and X-ray markings aligned but markings show to the right: flipped chamber with a 180° turn.
(2) In accordance with 2010 HAS Checklist, 2012 SF2H recommendations and NICE guidelines since 2002.
(3) Lapalu J & al., Totally Implantable Port Management: Impact of positive pressure during needle withdrawal on catheter ti  p occlusion (An experimental study), Journal of Vascular Access, 2010; 11 : 46-51.
(4) See instruction for use CD00675.
(5) CDC Guidelines for Basic Infection Control And Prevention Plan for Outpatient Oncology Settings, 2011.
** For micro references

CONTACT PROMECON GMBH

Admiralitätstraße 60 // 20459 Hamburg

Tel. : +49 40 3690 1690
Fax : +49 40 3690 1699
info@promecon-medical.de